Be sure to pack up any item well (we suggest you take it to a shipping store - towels, paper and used bubble wrap are NOT good packing materials) and ship them in to us by method of your choice (please allow up to 5 business days for shipping).  Be sure to include a note with who to contact about the unit and what it is or is not doing (error messages, etc).   All items are returned via UPS (again, allow up to 5 business days for shipping).
For more information, please contact us at 888-996-9968 9 a.m. to 5 p.m. Monday through Friday.
Service Tips: Anesthesia Maintenance Checklist
The following items should be checked on a regular basis.

O2 supply pressure

 

  • Pressure from H-tank or other O2 supply should be set at 50 PSI
  • Check that you have a backup supply at all times as well
Anesthetic agent level
  • Fill at the end of the day in a well ventilated area. Keep fill funnel clean.
Absorber

a) CO2 Absorbent replace every 10 hours of use -OR-
b) CO2 Absorbent replaced once a week (even if not used due to moisture content loss) -OR-
c) CO2 Absorbent replaced when 50% color change ONLY during first continuous use (not a good general rule of thumb)

Waste gas scavenging
  • Check that any charcoal canister is not over maximum weight
  • Check that canister is not against the floor or other surface where air cannot escape canister.
  • Check that exhaust system connected and working
  • Be sure recovery for Isoflurane patients is in well ventilated area.
  • Use only cuffed ET tubes, induction chamber or mask.
  • Use anesthetic only in well ventilated areas.
Breathing system test
  • Absorber leak test: No more than 200 ml (400 ml on old machines) O2 flow to hold 40cm H20
  • To perform test:  close pop-off, block O2 hose end and flush O2 to 40cm, if system does not hold pressure, there is a leak.
Pop off function
  • Ensure the pop-off valve or RSO restrictor tightens and loosens smoothly and functions properly.
Breathing bag and tubes
  • When back inflates for pop-off test, be sure bag is not leaking and both bag and tubes are in good condition
Flowmeter
  • Test flow meters to insure proper smooth function
Endotracheal tubes
  • Test for cuff leaks by inflating and inspecting cuffs.
Preventative maintenance
  • Do preventative maintenance on anesthesia machines every 6 months, have vaporizers fully serviced once per year.