The Baxter 6201 uses only standard Baxter solution administration sets with configurable settings allow the pump to be custom-tailored to best suit the hospitals needs. Computer control capability allows remote control or monitoring of the Flo-Gard 6201 pump via a standard RS232-C interface. The Baxter 6201 Pumps a wide variety of fluids, including blood and fat emulsions; suitable for use in epidural administration. The safety clamp automatically occludes the tubing when the pump door is opened. The Occlusion sensors detect both upstream and downstream restrictions. Sensitivity of the downstream occlusion sensor is selectable through the pumps configuration. Flow check display shows resistance to flow. Has an Ultrasonic air-in-line detector. Sensitivity of the air detector is selectable through the pumps configuration. The Programmed Delivery Profile (PDP) enables manual programming or ramped calculation of up to 10 sequential infusion programs for situations where multiple flow rates are indicated. The infusion pump slide clamp loading option provides flow shutoff when I.V. set is removed from the pump. Locking control panel helps prevent patient tampering. There is an independent secondary medication program. The pump automatically switches over to the primary program upon completion. Volume-Time programming automatically calculates flow rate. Automatic self-test routine checks for proper function before use. Five-hour memory retains infusion data after power-off. Easily replaceable fuse, battery, and power cord.
Features:
- Keep Vein Open (KVO) Rate 5 mL/hr or programmed rate, whichever is less
Battery 12 Volt, 2.0 Ah sealed lead-acid
- Battery Life Approximately 6 hours with pump running at a rate from 1 to 1400 mL/hr
- Battery Recharge 8 hours for more than 80% recharge
- AC Power Requirements 110/120V, 60 Hz
- Power Cord 2.9 m (9 ft) long, with hospital grade plug
- Fuse 0.5 A, 250V, SB, 5.2 mm (13/64 in) x 20 mm (25/32 in)
- Leakage Current Less than 50 microamps (using UL-544 specified test methods)
- Weight Approximately 5.3 kg (11.6 lbs)
- Dimensions 20 cm W x 13 cm D x 29 cm H (7.9″ W x 5.1″ D x 11.4″ H)
- Flow Rate Range Primary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-1999 mL/hr in 1 mL increments. Upper limit can be reduced by authorized service personnel.
- Secondary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-999 mL/hr in 1 mL increments
- VTBI Range 1.0-99.9 mL in 0.1 mL increments or 1-9999 mL in 1 mL increments for both primary and secondary. Upper limit can be reduced by authorized service personnel.
- Ramp PDP Parameter Ranges Total VTBI: 100-9,999 in 1 mL increments. Upper limit can be reduced by authorized service personnel.
- Delay Time: 00:00 – 08:00 hours in minute increments.
- Run Time: 01:00 – 16:00 hours in minute increments.
- Ramp Up: 00:00 to 45% of the Run Time, in minute increments.
- Ramp Down: 00-00 to the Run Time minus Ramp Up, in minute Increments.
- Air-in-Line Detection Factory set to NORM, which causes the pump to alarm on air bubbles approximately 75 uL or larger. The MIN setting causes the pump to alarm on air bubbles approximately 50 ul or larger.